Statutory Documents - IMO Publications and Documents - International Codes - NOx Technical Code (2008) - Technical Code on Control of Emission of Nitrogen Oxides from Marine Diesel Engines - Appendix 4 - Calibration of the analytical and measurement instruments - 5 Calibration procedure - 5.4 Calibration

5.4 Calibration

  5.4.1 Each normally used operating range shall be calibrated. Analysers shall be calibrated not more than 3 months before being used for testing or whenever a system repair or change is made that can influence calibration, or as per provided for by 1.3.2.2.

  5.4.2 Using purified synthetic air (or nitrogen) the CO, CO2, NOx and O2 analysers shall be set at zero. The (H)FID analyser shall be set to zero using purified synthetic air.

  5.4.3 The appropriate calibration gases shall be introduced to the analysers, the values recorded, and the calibration curve established accordingly.


Copyright 2022 Clasifications Register Group Limited, International Maritime Organization, International Labour Organization or Maritime and Coastguard Agency. All rights reserved. Clasifications Register Group Limited, its affiliates and subsidiaries and their respective officers, employees or agents are, individually and collectively, referred to in this clause as 'Clasifications Register'. Clasifications Register assumes no responsibility and shall not be liable to any person for any loss, damage or expense caused by reliance on the information or advice in this document or howsoever provided, unless that person has signed a contract with the relevant Clasifications Register entity for the provision of this information or advice and in that case any responsibility or liability is exclusively on the terms and conditions set out in that contract.