9.1 CO analyser interference check
Clasification Society 2024 - Version 9.40
Statutory Documents - IMO Publications and Documents - International Codes - NOx Technical Code (2008) - Technical Code on Control of Emission of Nitrogen Oxides from Marine Diesel Engines - Appendix 4 - Calibration of the analytical and measurement instruments - 9 Interference effects with CO, CO2, NOx and O2 analysers - 9.1 CO analyser interference check

9.1 CO analyser interference check

 Water and CO2 can interfere with the CO analyser performance. Therefore, a CO2 span gas having a concentration of 80% to 100% of full scale of the maximum operating range used during testing shall be bubbled through water at room temperature and the analyser response recorded. The analyser response must not be more than 1% of full scale for ranges equal to or above 300 ppm or more than 3 ppm for ranges below 300 ppm.


Copyright 2022 Clasifications Register Group Limited, International Maritime Organization, International Labour Organization or Maritime and Coastguard Agency. All rights reserved. Clasifications Register Group Limited, its affiliates and subsidiaries and their respective officers, employees or agents are, individually and collectively, referred to in this clause as 'Clasifications Register'. Clasifications Register assumes no responsibility and shall not be liable to any person for any loss, damage or expense caused by reliance on the information or advice in this document or howsoever provided, unless that person has signed a contract with the relevant Clasifications Register entity for the provision of this information or advice and in that case any responsibility or liability is exclusively on the terms and conditions set out in that contract.