8.3.1 The oxygen interference check shall be determined
when introducing an analyser into service and after major service
intervals.
8.3.2 A range shall be chosen where the oxygen
interference check gases will fall in the upper 50%. The test shall
be conducted with the oven temperature set as required. The oxygen
interference gases are specified in 2.2.4.
-
.1 The analyser shall be zeroed.
-
.2 The analyser shall be spanned with the 21%
oxygen blend.
-
.3 The zero response shall be rechecked. If it
has changed more than 0.5% of full scale (FS) steps 8.3.2.1 and 8.3.2.2
shall be repeated.
-
.4 The 5% and 10% oxygen interference check gases
shall be introduced.
-
.5 The zero response shall be rechecked. If it
has changed more than ± 1% of full scale, the test shall be
repeated.
-
.6 The oxygen interference (%O
2
I) shall be calculated for each mixture in step .4 as follows:
-
where: analyser response is (A/% FS at A)
· (%FS at B)
-
where:
-
.7 The % of oxygen interference (%O
2
I) shall be less than ± 3.0% for all required
oxygen interference check gases prior to testing.
-
.8 If the oxygen interference is greater than
± 3.0%, the air flow above and below the manufacturer’s
specifications shall be incrementally adjusted, repeating 8.1 for
each flow.
-
.9 If the oxygen interference is greater than
± 3.0% after adjusting the air flow, the fuel flow, and thereafter
the sample flow shall be varied, repeating 8.1 for each new setting.
-
.10 If the oxygen interference is still greater than
± 3.0%, the analyser, (H)FID fuel, or burner air shall be repaired or replaced
prior to testing. This clause shall then be repeated with the repaired or replaced
equipment or gases.