2.2.1 Mixtures of gases having the following chemical
compositions shall be available:
-
.1 CO and purified nitrogen;
-
.2 NOx and purified nitrogen the amount
of NO2 contained in this calibration gas shall not exceed
5% of the NO content);
-
.3 O2 and purified nitrogen;
-
.4 CO2 and purified nitrogen; and
-
.5 CH4 and purified synthetic air or
C3H8 and purified synthetic air.
-
Note:
Other gas combinations are
allowed provided the gases do not react with one another.
2.2.2 The true concentration of a calibration
and span gas must be within ± 2% of the nominal value. All
concentrations of calibration and span gases shall be given on a volume
basis (volume per cent or volume ppm).
2.2.3 The gases used for calibration and span
may also be obtained by means of precision blending devices (gas dividers),
diluting with purified N2 or with purified synthetic air.
The accuracy of the mixing device must be such that the concentration
of the blended calibration gases is accurate to within ±2%.
This accuracy implies that primary gases used for blending must be
known to an accuracy of at least ±1%, traceable to national
or international gas standards. The verification shall be performed
at between 15 and 50% of full scale for each calibration incorporating
a blending device. Optionally, the blending device may be checked
with an instrument which by nature is linear, e.g., using NO gas with
a CLD. The span value of the instrument shall be adjusted with the
span gas directly connected to the instrument. The blending device
shall be checked at the used settings and the nominal value shall
be compared to the measured concentration of the instrument. This
difference shall in each point be within ±1% of the nominal
value. This linearity check of the gas divider shall not be performed
with a gas analyser which was previously linearized with the same
gas divider.
2.2.4 Oxygen interference check gases shall contain propane
or methane with 350 ppmC ± 75 ppmC hydrocarbon. The concentration shall be determined to
calibration gas tolerances by chromatographic analysis of total hydrocarbons plus
impurities or by dynamic blending. Nitrogen shall be the predominant diluent with the
balance oxygen. Blends required are listed in table 5.
Table 5 Oxygen interference check
gases
O2
concentration
|
Balance
|
21 (20 to 22)
|
Nitrogen
|
10 (9 to 11)
|
Nitrogen
|
5 (4 to 6)
|
Nitrogen
|