1.1 The components included in an exhaust gas
analysis system for the determination of the concentrations of CO,
CO2, NOx, HC and O2 are shown in
figure 1. All components in the sampling gas path must be maintained
at the temperatures specified for the respective systems.
Figure 1 Arrangement of exhaust gas analysis system
1.2 An exhaust gas analysis system shall include
the following components. In accordance with chapter
5 of this Code equivalent arrangements and components may,
subject to approval by the Administration, be accepted.
-
.1 SP Raw exhaust gas sampling probe
-
A stainless steel, straight, closed-end, multi-hole
probe. The inside diameter shall not be greater than the inside diameter
of the sampling line. The wall thickness of the probe should not be
greater than 1 mm. There should be a minimum of three holes in three
different radial planes sized to sample approximately the same flow.
-
For the raw exhaust gas, the sample for all components
may be taken with one sampling probe or with two sampling probes located
in close proximity and internally split to the different analysers.
-
Note:
If exhaust pulsations or engine
vibrations are likely to affect the sampling probe, the wall thickness
of the probe may be enlarged subject to the approval of the Administration.
-
.2 HSL1 Heated sampling line
-
The sampling line provides a gas sample from a single probe
to the split point(s) and the HC analyser. The sampling line shall
be made of stainless steel or polytetrafluoroethylene (PTFE) and have
a 4 mm minimum and a 13.5 mm maximum inside diameter.
-
The exhaust gas temperature at the sampling probe shall
not be less than 190oC. The temperature of the exhaust
gas from the sampling point to the analyser shall be maintained by
using a heated filter and a heated transfer line with a wall temperature
of 190oC ± 10oC.
-
If the temperature of the exhaust gas at the sampling probe
is above 190oC, a wall temperature greater than 180oC
shall be maintained.
-
Immediately before the heated filter and the HC analyser
a gas temperature of 190oC ± 10oC shall
be maintained.
-
.3 HSL2 Heated NOx sampling
line
-
The sampling line shall be made of stainless steel or PTFE
and maintain a wall temperature of 55oC to 200oC,
up to the converter C when using a cooling unit B, and up to the analyser
when a cooling unit B is not used.
-
.4 HF1 Heated pre-filter (optional)
-
The required temperature shall be the same as for HSL1.
-
.5 HF2 Heated filter
-
The filter shall extract any solid particles from the gas
sample before the analyser. The temperature shall be the same as for
HSL1. The filter shall be changed as necessary.
-
.6 HP Heated sampling pump (optional)
-
The pump shall be heated to the temperature of HSL1
-
.7 SL Sampling line for CO, CO2 and
O2
-
The line shall be made of PTFE or stainless steel. It may
be heated or unheated.
-
.8 CO2/CO Carbon dioxide and
carbon monoxide analysers
-
Non-dispersive infrared (NDIR) absorption. Either separate
analysers or two functions incorporated into a single analyser unit.
-
.9 HC Hydrocarbon analyser
-
Heated flame ionization detector (HFID). The temperature
shall be kept at 180oC to 200oC.
-
.10 NOx Nitrogen oxides analyser
-
Chemiluminescent detector (CLD) or heated chemiluminescent
detector (HCLD). If a HCLD is used, it shall be kept at a temperature
of 55oC to 200oC.
-
.11 C converter
-
A converter shall be used for the catalytic reduction of
NO2 to NO prior to analysis in the CLD or HCLD.
-
.12 O2 Oxygen analyser
-
Paramagnetic detector (PMD), zirconium dioxide (ZRDO) or
electrochemical sensor (ECS). ZRDO shall not be used for dual fuel or gas-fuelled
engines.
-
Note:
In the arrangement shown O2 is
measured on a dry basis. O2 may also be measured on a wet
basis in which case the analyser shall be of the ZRDO type.
-
.13 B cooling unit
-
To cool and condense water from the exhaust sample. The
cooler shall be maintained at a temperature of 0oC to 4oC
by ice or refrigerator. If water is removed by condensation, the sample
gas temperature or dew point shall be monitored either within the
water trap or downstream. The sample gas temperature or dew point
shall not exceed 7oC.
1.3 The analysers shall have a measuring range
appropriate for the accuracy required to measure the concentrations
of the exhaust gas components (see 1.6) and 5.9.7.1 of this Code. It is recommended that the analysers be operated
such that the measured concentration falls between 15% and 100% of
full scale, where full scale refers to the measurement range used.
1.4 If the full-scale value is 155 ppm (or ppmC)
or less, or if read-out systems (computers, data loggers) that provide
sufficient accuracy and resolution below 15% of full scale are used,
concentrations below 15% of full scale are also acceptable. In this
case, additional calibrations are to be made to ensure the accuracy
of the calibration curves.
1.5 The electromagnetic compatibility (EMC) of
the equipment shall be such as to minimize additional errors.
1.6 Accuracy
1.6.1 Definitions
ISO 5725-1: 1994/Cor 1: 1998, Accuracy (trueness and precision) of
measurement methods and results Part 1: General principles and definitions, Technical
Corrigendum 1.
ISO 5725-2: 1994, Accuracy (trueness and precision) of measurement methods
and results Part 2: Basic method for the determination of repeatability and
reproducibility of a standard measurement method.
1.6.2 An analyser shall not deviate from the nominal
calibration point by more than ± 2% of the reading over the whole measurement range
except zero, or ± 0.3% of full scale, whichever is larger. The accuracy shall be
determined according to the calibration requirements laid down in section 5 of appendix
4 of this Code.
1.7 Precision
The precision, defined as 2.5 times the standard deviation of 10 repetitive
responses to a given calibration or span gas, shall be not greater than ± 1% of
full-scale concentration for each range used above 100 ppm (or ppmC) or ± 2% of each
range used below 100 ppm (or ppmC).
1.8 Noise
The analyser peak-to-peak response to zero and calibration or span gases
over any 10second period shall not exceed 2% of full scale on all ranges used.
1.9 Zero drift
Zero response is defined as the mean response, including noise, to a zero
gas during a 30second time interval. The drift of the zero response during a one-hour
period shall be less than 2% of full scale on the lowest range used.
1.10 Span drift
Span response is defined as the mean response, including noise, to a span
gas during a 30-second time interval. The drift of the span response during a one-hour
period shall be less than 2% of full scale on the lowest range used.