3.1.1 Before and after temperature conditioning
in accordance with 3.1.2, the foam concentrate should show no visual
sign of stratification, non-homogeneity or sedimentation.
3.1.2 Freezing and thawing test
-
.1 Apparatus:
-
.1 freezing chamber, capable of achieving temperatures
required as stated in .2.1 below;
-
.2 polyethylene tube, approximately 10 mm diameter,
400 m long and sealed and weighted at one end, with suitable spacers
attached (Figure 1 of MSC/Circ.582 shows a typical form); and
-
.3 500 ml cylinder approximately 400 mm high and
65 mm in diameter.
-
.2 Procedure:
-
.1 set the temperature of the freezing chamber
to a temperature which is 10°C below the freezing point of the
sample measured in accordance with BS 5117: Section 1.3 (excluding
5.2 in the standard);
To prevent the glass measuring cylinder
from breaking, due to expansion of the foam concentrate on freezing,
insert the tube into the measuring cylinder, sealed end downward,
weighted if necessary to void flotation, the spacers ensuring it remains
approximately on the central axis of the cylinder.
Place the sample in the cylinder in the chest, cool and maintain
at the required temperature for 24 h. At the end of this period thaw
the sample for not less than 24 h and not more than 96 h in an ambient
temperature range of 20°C to 25°C.
-
.2 repeat .2.1 above
three times to give four cycles of freezing and thawing;
-
.3 examine the sample for stratification and non-homogeneity;
and
-
.4 condition the sample for 7 days at 60°C
followed by one day at room temperature.