The nitrogen oxides analyser shall be of the chemiluminescent
detector (CLD) or heated chemiluminescent detector (HCLD) type with
a NO2/NO converter, if measured on a dry basis. If measured
on a wet basis, a HCLD with converter maintained above 55oC
shall be used, provided the water quench check (see section 9.2.2
of appendix 4 of this Code) is satisfied.
For both CLD and HCLD, the sampling path shall be maintained at a
wall temperature of 55oC to 200oC up to the
converter for dry measurement, and up to the analyser for wet measurement.