1.3.1 Pure and calibration gases, as required,
should comply with the NOx Technical Code, appendix 4, paragraphs 2.1 and 2.2. Declared concentrations
should be traceable to national and/or international standards. Calibration
gases should be in accordance with the analysing equipment manufacturer's
recommendations.
1.3.2 Analyser span gases should be between 80%-100%
of the analyser scale being spanned.
Note Under certain conditions, only a span gas for the NOx
analyser could be necessary for calibration of the analytical instruments.
A NO span gas for a NOx analyser can be applied as a zero gas for
an O2 or CO2 analyser if that span gas is balanced
only with nitrogen. Ambient air may be applied both as a span gas
for an O2 analyser (i.e. 20.9% O2) and as a
zero gas for a NOx analyser, provided that the ambient air is uncontaminated
by exhaust gas.