5.4 Calibration

  5.4.1 Each normally used operating range shall be calibrated.

  5.4.2 Using purified synthetic air (or nitrogen), the CO, CO2, NOx and O2 analysers shall be set at zero.

  5.4.3 The appropriate calibration gases shall be introduced to the analysers, the value recorded, and the calibration curve established according to 5.5 below.

  5.4.4 The zero setting shall be rechecked and the calibration procedure repeated, if necessary.


Copyright 2022 Clasifications Register Group Limited, International Maritime Organization, International Labour Organization or Maritime and Coastguard Agency. All rights reserved. Clasifications Register Group Limited, its affiliates and subsidiaries and their respective officers, employees or agents are, individually and collectively, referred to in this clause as 'Clasifications Register'. Clasifications Register assumes no responsibility and shall not be liable to any person for any loss, damage or expense caused by reliance on the information or advice in this document or howsoever provided, unless that person has signed a contract with the relevant Clasifications Register entity for the provision of this information or advice and in that case any responsibility or liability is exclusively on the terms and conditions set out in that contract.