5.5.1.1 The analyser calibration curve shall be
established by at least five calibration points (excluding zero) spaced
as uniformly as possible. The highest nominal concentration shall
be greater than or equal to 90% of full scale.
5.5.1.2 The calibration curve is calculated by
the method of least squares. If the resulting polynomial degree is
greater than 3, the number of calibration points (zero included) shall
be at least equal to this polynomial degree plus 2.
5.5.1.3 The calibration curve shall not differ
by more than ±2% from the nominal value of each calibration
point and by more than ±1% of full scale at zero.
5.5.1.4 From the calibration curve and the calibration
points, it is possible to verify that the calibration has been carried
out correctly. The different characteristic parameters of the analyser
shall be indicated, particularly: