2.1 The MARPOL sample should be delivered by the
competent authority to the laboratory.
2.2 The laboratory should:
-
.1 record the details of the seal number and the
sample label on the test record;
-
.2 confirm that the condition of the seal on the
MARPOL sample has not been broken; and
-
.3 reject any MARPOL sample where the seal has
been broken.
2.3 If the seal of the MARPOL sample has not been
broken, the laboratory should proceed with the verification procedure
and should:
-
.1 ensure that the MARPOL sample is thoroughly
homogenized;
-
.2 draw two sub-samples from the MARPOL sample;
and
-
.3 reseal the MARPOL sample and record the new
reseal details on the test record.
2.4 The two sub-samples should be tested in succession,
in accordance with the specified test method referred to in Appendix V. For the purposes of this
verification procedure, the results of the test analysis should be
referred to as Aand B.
-
.1 If the results of Aand B
are within the repeatability (r) of the test method, the results should
be considered valid.
-
.2 If the results of Aand B
are not within the repeatability (r) of the test method, both results
should be rejected and two new sub-samples should be taken by the
laboratory and analysed. The sample bottle should be resealed in accordance
with paragraph 2.3.3 above after the new sub-samples have been taken.
2.5 If the test results of Aand
B are valid, an average of these two results should
be calculated thus giving the result referred to as X.
-
.1 If the result of X is equal to
or falls below the standards required by Annex
VI, the fuel oil should be deemed to meet the requirements.
-
.2 If the result of X is greater
than the standards required by Annex VI,
Verification Procedure Stage 2 should be conducted; however, if the
result of X is greater than the specification limit
by 0.59R (where R is the reproducibility of the test method), the
fuel oil should be considered non-compliant and no further testing
is necessarily.