1.3.1 The calibration of all measuring instruments
shall comply with the requirements as set out in tables 1, 2, 3 and
4 and shall be traceable to standards recognized by the Administration.
Additional engine measurements may be required by the Administration,
and such additional measuring instruments used shall comply with the
appropriate deviation standard and calibration validity period.
1.3.2 The instruments shall be calibrated:
-
.1 in time intervals not greater than as given
in tables 1, 2, 3 and 4; or
-
.2 in accordance with alternative calibration
procedures and validity periods subject to such proposals being submitted
in advance of the tests and approved by the Administration.
Note:
The deviations given in tables
1, 2, 3, and 4 refer to the final recorded value, which is inclusive
of the data acquisition system.
Table 1 Permissible deviations and
calibration validity periods of instruments for engine-related parameters for
measurements on a test bed
No.
|
Measurement instrument
|
Permissible deviation
|
Calibration validity
period (months)
|
1
|
Engine speed
|
± 2% of reading or ± 1% of engine’s
maximum value, whichever is larger
|
3
|
2
|
Torque
|
± 2% of reading or ± 1% of engine’s
maximum value, whichever is larger
|
3
|
3
|
Power (where measured
directly)
|
± 2% of reading or ± 1%
of engine’s maximum value, whichever is larger
|
3
|
4
|
Fuel consumption
|
± 2% of engine’s maximum
value
|
6
|
5
|
Air consumption
|
± 2% of reading or ± 1%
of engine’s maximum value, whichever is larger
|
6
|
6
|
Exhaust gas flow
|
± 2.5% of reading or ±
1.5% of engine’s maximum value, whichever is larger
|
6
|
Table 2 Permissible deviations and
calibration intervals periods of instruments for other essential parameters for
measurements on a test bed
No.
|
Measurement instrument
|
Permissible deviation
|
Calibration validity
period (months)
|
1
|
Temperatures ≤ 327oC
|
± 2oC absolute
|
3
|
2
|
Temperatures > 327oC
|
± 1% of reading
|
3
|
3
|
Exhaust gas pressure
|
± 0.2 kPa absolute
|
3
|
4
|
Charge air pressure
|
± 0.3 kPa absolute
|
3
|
5
|
Atmospheric pressure
|
± 0.1 kPa absolute
|
3
|
6
|
Other pressures ≤ 1000 kPa
|
± 20 kPa absolute
|
3
|
7
|
Other pressures > 1000 kPa
|
± 2% of reading
|
3
|
8
|
Relative humidity
|
± 3% absolute
|
1
|
Table 3 Permissible deviations and
calibration validity periods of instruments for engine-related parameters for
measurements on board a ship when the engine is already pre-certified
No.
|
Measurement
instrument
|
Permissible
deviation
|
Calibration validity
period (months)
|
1
|
Engine speed
|
± 2% of engine’s
maximum value
|
12
|
2
|
Torque
|
± 5% of engine’s
maximum value
|
12
|
3
|
Power (where measured
directly)
|
± 5% of engine’s
maximum value
|
12
|
4
|
Fuel consumption
|
± 4% of engine’s
maximum value
|
12
|
5
|
Air consumption
|
± 5% of engine’s
maximum value
|
12
|
6
|
Exhaust gas flow
|
± 5% of engine’s
maximum value
|
12
|
Table 4 Permissible deviations
calibration validity period of instruments for other essential parameters for
measurements on board a ship when the engine is already pre-certified
No.
|
Measurement
instrument
|
Permissible
deviation
|
Calibration validity
period (months)
|
1
|
Temperatures ≤ 327oC
|
± 2oC
absolute
|
12
|
2
|
Temperatures > 327oC
|
± 15oC
absolute
|
12
|
3
|
Exhaust gas pressure
|
± 5% of engine’s
maximum value
|
12
|
4
|
Charge air pressure
|
± 5% of engine’s
maximum value
|
12
|
5
|
Atmospheric pressure
|
± 0.5% of
reading
|
12
|
6
|
Other pressures
|
± 5 % of
reading
|
12
|
7
|
Relative humidity
|
± 3% absolute
|
6
|