Each normally used operating range shall be checked prior
to each analysis in accordance with the following procedure:
-
.1 the calibration shall be checked by using a
zero gas and a span gas whose nominal value shall be more than 80%
of full scale of the measuring range; and
-
.2 if, for the two points considered, the value found
does not differ by more than ± 4% of full scale from the declared reference value,
the adjustment parameters may be modified. If this is not the case, a new
calibration curve shall be established in accordance with 5.5 above.